LONG TERM STABILITY IN CW-CRDS EXPERIMENTS.

HAIFENG HUANG, Department of Chemistry, University of Virginia, Charlottesville VA, 22904-4319; KEVIN K. LEHMANN, .

Allan variance can be used to characterize the slow drift of CW cavity ring-down spectroscopy (CW-CRDS) system, caused by the instability of experimental environments. By including the differential measurement of on- and off-peak decay rates in the system, the drifting of both decay rates can cancel out greatly. A sensitivity of 5.6 × 10-12 cm-1 during an optimum integration time of 15.4 minutes has been obtained with our CW-CRDS system. This sensitivity corresponds a 3 sigma methane detection limit of 0.3 parts per billion by volume (ppbv) in N2 at 20 torr or 37 parts per trillion by volume (pptv) in N2 at 760 torr measurement pressure, near 1.65 µm laser wavelength.